Overview
This section introduces sample measurement procedures, with Part 15 covering procedures for ICP-OES and later parts covering measurement for ICP-MS. This section is for advanced chemists who are already familiar with using the instrumentation. Most of the information and tips I supply here will not be found in a standard textbook; rather, they reflect the major considerations of sample measurement based on my own personal experience.
When developing an ICP-OES measurement procedure, the following concerns should be addressed:
- Line selection
- Sensitivity requirements
- Precision requirements
- Spectral interferences and background correction
- Matrix effects
- Method validation (to be addressed in a later part of this section)
Line Selection
The first step in line selection is to simply choose the line(s) that meet the sensitivity requirements for your measurement. More than one line may be necessary due to spectral interferences.
Sensitivity
Make sure that the line is not too sensitive (i.e., going outside the linear working range). Dilutions are easy, and hopefully you'll have more than one line that meets your lower limit of detection and quantitative measurement requirements. If you do not have line tables, we have listed the three most popular lines for each element (also included are detection limits for radial and axial view, as well as major interferences) in our Interactive Periodic Table.
Precision
The precision of an ICP measurement is a function of many factors that are beyond our control. However, if you require an improvement in precision, consider the following:
- Try to keep the analyte concentration well within the linear working range. When the concentration falls <100 times the detection limit, the precision begins to fall.
- Avoid lines requiring spectral correction.
- Avoid lines that are in spectrally complex regions requiring sophisticated background correction algorithms.
- Increasing the integration time to as high as 5 seconds should improve the precision. It should not be increased any higher.
- Use an all-glass introduction system, including a glass concentric nebulizer. Your washout times will improve along with your precision.
- Be wary of salting-out effects when analyzing high-salt-containing solutions, but do not use high solids nebulizers unless necessary.
- If your samples and standards have the same background, then you should skip background correction.
- If the matrices of the samples and standards can be matched, then eliminate the peristaltic pump and go "free flow." The pump introduces pulsing, and the pump tubing stretches, causing the introduction rate to gradually change. If you can avoid the pump, then make sure the liquid levels of your samples and standards are the same.
NOTE: Slight differences in hydrostatic pressure will make a difference.
- Do not begin the measurement unless the instrument has been allowed to warm up for at least an hour. Additionally, working in a temperature-controlled atmosphere is very advantageous.
- Clean your introduction system and replace the tubing when (a) you notice an increase in precision when using glass introduction systems, (b) you notice a "bend or crink" in the Teflon tubing, (c) the torch begins to build up residue.
Spectral Interferences
The spectral tables are very useful, but they are no replacement for a spectral interference study. These studies should be performed on every instrument when it is installed, and at least annually thereafter. These time-consuming and tedious studies are very necessary. They are performed by aspirating 1000 µg/mL solutions of the potential interfering element and looking at the spectral regions around possible analyte lines for unwanted spectral interferences. The types of spectral interferences encountered are direct spectral overlap, wing overlap, and near neighbors that may cause background correction problems. The spectral region for the analyte, when a blank solution or the possible interfering element solution is analyzed, is hopefully the same, indicating no spectral interference.
A trace metals analysis of the 1000 µg/mL solutions helps distinguish between a true direct spectral overlap and the presence of the analyte in the interfering element solution as an impurity. The absence of impurity data requires confirmation of the suspected direct spectral overlap using an alternate analyte line or another technique, such as flame atomic absorption. Therefore, it is important to use standards with accurate trace metals impurity data.