Measurement and Results - Using a cesium (Cs) buffer for ICP-OES measurements

 
I will be needing to perform Na and K analysis on an axial ICP-OES and see that you sell a cesium ionization buffer that is 1% cesium (CSN-ISB-500ML) and 5% cesium. At what concentration should the cesium be in my test samples?
Sort By    Oldest  |  Newest
Responses
admin Total posts: 529

Our standard procedure is to add Cs at a concentration of 500 ppm to all test solutions (standards and samples) measured by axial-view ICP-OES, and we have had good results following this approach.

Posted: 09/05/18 20:59:37