Measurement and Results - Replacing yttrium (Y) internal standard for ICP-MS

 
I am looking for a quote on a new Internal Standard for our ICP-MS. We currently use a custom internal standard provided by IV, but need to replace the Y in the standard with another metal. We are looking for suggestions you may have. Thank you for your time.
Sort By    Oldest  |  Newest
Responses
admin Total posts: 529

We would recommend either gallium (Ga) or rhodium (Rh) for that mass range.  We are able to provide both of those elements in HNO3, free of any chloride or fluoride which would greatly reduce the risk of introducing potentially problematic chemical interferences in your analyses.  Please let us know if these elements would not work for some reason, and we will make other recommendations.

Posted: 09/04/18 22:08:14