Cart

Sample Preparation - Poor Mn recovery for soil standards

 
I have recently encountered an odd issue with Mn recovery from solid samples. Basically, recovery of Mn from Metals in Soil standards from Sigma Aldrich have a roughly 1 in 4 chance of being outside of Sigma’s limits. This can be either high or low. I have been using CEM’s MARS microwave, carousel, and tubes. Following EPA 3051 for the digestion, then analysis via ICP-OES. This odd recovery seems to be happening regardless of sample (have tested several lot #s) or location in the carousel. I’m hoping that you would have some experience with this problem, or perhaps have some suggestions. I have spoken with CEM support, and they are unable to provide any explanation or solution.
Sort By    Oldest  |  Newest
Responses
admin Total posts: 529

There are a number of possibilities to explain your results, but we will assume that sample inhomogeneity and/or poor characterization of the soil standards are not the cause. These assumptions are reasonable for standards, but ultimately may not be appropriate in the case of your high results. The first consideration is that 3051 is not intended as a total sample decomposition method, so are you seeking to quantify only extractable metals? With that in mind, what other metals are you measuring for, are you using only HNO3 or the general 3:1 ratio of HNO3 to HCl, and do you observe clear solutions free of particulates following digestion? If some sample material remains, so you filter or centrifuge prior to analysis? If the soil standards are homogenous and accurate with respect to their stated values, then the sample preparation is the most likely cause for your results, and modifying details of that method might help, e.g., greater acid-to-sample ratio, different acid combinations, or longer digestion times. If a single digested sample does not give repeatable results over several different ICP measurements, then the analysis itself is likely part of the problem. It might be helpful to start there; can a single digested sample generate the same ICP results over multiple runs over several days. These are general considerations as there are many potential causes for your unusual Mn results. If you could provide any additional information, it’s possible we could offer more specific recommendations.

Posted: 09/06/18 04:16:49

1 Item

Show per page