For ICP-MS or ICP-OES why are internal standards required? How do you decide the concentration of the internal standard, and is there any relation between the internal standard and sample?
Sort By    Oldest  |  Newest
Responses
admin Total posts: 529

Internal standards are not required for ICP-OES or ICP-MS measurements. However, they can significantly improve the accuracy of the analytical data, as an internal standard provides a reference that can be used to correct for variability between the calibration standards and the samples. The most common example is correcting for matrix differences between the calibration standards and samples, and the key criteria for using an internal standard are: 1) it must not be present in the sample, 2) it must be free from interferences, 3) it’s added in equal concentrations to all analytical solutions (blanks, calibration standards, and samples). The ICP Operations Guide on our website discusses these considerations in more detail (https://www.inorganicventures.com/standard-addition-internal-standardization-and-isotope-dilution) and may provide useful.

Posted: 09/04/18 22:19:12