OverviewThis section introduces sample measurement procedures, with part 15 covering procedures for ICP-OES and later parts covering measurement for ICP-MS. This section is for advanced chemists who are already familiar with using the instrumentation. Most of the information and tips I supply here will not be found in a standard textbook; rather they reflect the major considerations of sample measurement based on my own personal experience. When developing an ICP-OES measurement procedure, the following concerns should be addressed:
Line SelectionThe first step in line selection is to simply choose the line(s) that meet the sensitivity requirements for your measurement. More than one line may be necessary due to spectral interferences. SensitivityMake sure that the line is not too sensitive (i.e. - going outside the linear working range). Dilutions are easy and hopefully you'll have more than one line that meets your lower limit of detection and quantitative measurement requirements. If you do not have line tables, we have listed the three most popular lines for each element (also included are detection limits for radial and axial view, as well as major interferences) in our Interactive Periodic Table. PrecisionThe precision of an ICP measurement is a function of many factors that are beyond our control. However, if you require an improvement in precision, consider the following:
Spectral InterferencesThe spectral tables are very useful, but they are no replacement for a spectral interference study. These studies should be performed on every instrument when it is installed and at least annually thereafter. These time-consuming and tedious studies are very necessary. They are performed by aspirating 1000 µg/mL solutions of the potential interfering element and looking at the spectral regions around possible analyte lines for unwanted spectral interferences. The types of spectral interferences encountered are direct spectral overlap, wing overlap, and near neighbors that may cause background correction problems. The spectral region for the analyte when a blank solution or the possible interfering element solution are analyzed are hopefully the same, indicating no spectral interference. A trace metals analysis of the 1000 µg/mL solutions helps distinguish between a true direct spectral overlap and the presence of the analyte in the interfering element solution as an impurity. The absence of impurity data requires confirmation of the suspected direct spectral overlap using an alternate analyte line or another technique such as flame atomic absorption. Therefore, it is important to use standards with accurate trace metals impurity data. Fig. 15.1
Spectrum of a high Ca containing matrix (red line) causing a sloping background for the Cu 219.959 or Ge 219.871 lines are used. Background correction is difficult at best in these situations. ![]() Fig. 15.2
Spectrum of a high concentration of Fe (red line) showing a direct spectral overlap upon the B 208.892 line and a wing overlap on a B 208.959. ![]() Fig. 15.3
The following shows a wing overlap interference of Fe (red line) upon the Ba 233.527 nm line. ![]() Fig. 15.4
Have you ever wondered how you can get negative numbers? It is likely due to a spectral situation similar to the one shown below. A nearby Fe line interferes with the background correction upon the Al 396.152 nm line. ![]() Example of a MethodThe following method (MEB) covers a wide range of elements. This method is currently under development and is presented for line selection purposes - it is not a recommended method. The lines shown were chosen on the basis of sensitivity and relative freedom from spectral interference. It is important to point out that this method is intended for an axial view instrument. Cs is used as an internal standard / buffer (double duty) in this method. Data collected over the past 8 months indicates the method is giving accurate and reliable data. However, at least another 12 months of work is planned before releasing this method due to the unorthodox use of Cs as an internal standard. We believe this is because the level of Cs used 'overwhelms' the matrix. Matrix EffectsMatrix effects are arguably the most subtle danger to the ICP-OES analyst. Slight differences in the matrix can cause a considerable systematic error having a definite bias. Internal standardization has been the most common approach toward correcting this problem and this approach does work well in many cases. The choice of the internal standard is critical. Please consider the following questions before using the internal standard approach for matrix correction:
The technique of standard additions is a tedious but more reliable approach for matrix correction. For unknown matrices it may well be the fastest approach. When using standard additions on unknown matrices, it is possible to have severe spectral and background correction problems. It is cautioned here that at least two spectral lines should be used and the spectral region carefully scanned and studied. Attempt to make the additions approximately 2x and 3x, where x is the analyte concentration. Further Reading |
|