Key Instrument ParametersICP Operations Guide: Part 9The performance characteristics of an ICP is a function of a variety of instrumental parameters. Current instrumentation has many parameters that are fixed by the manufacturer and all instrumentation will come with recommended settings for those parameters that are not. The purpose of this section is to point out the key parameters that will require adjustment on a regular basis. This discussion will be limited to the introduction of the analyte as a nebulized solution and Ar as the plasma gas. Gas Flow RatesThere are three gas flow rates for the common torch designs. The outer gas flow is sometimes referred to as the coolant or plasma gas flow; the middle or intermediate gas flow is sometimes referred to as the auxiliary gas flow; and the central gas flow is referred to as the sample or nebulizer gas flow. When working with aqueous solutions, the outer and intermediate gas flows do not have a great impact upon the performance characteristics and the values suggested by the manufacturer should be used for common applications. However, the sample gas flow rate will vary between nebulizers of the same design and require adjustment on a regular basis. Sample Ar Gas Flow for ICP-OESAssuming sample solution is not significantly limited, the main consideration when adjusting the sample Ar gas flow is that of precision. Increasing the sample Ar gas flow does not necessarily increase the emission intensity. The objective in setting this flow rate is to obtain the best detection limit. Noisy signals will typically result from higher flow rates that will serve to degrade the stability of the plasma, increase the short-term measurement precision and consequently give poorer detection limits. The following considerations should prove helpful:
Applied Power for ICP-OESThe second key parameter that the operator may wish to vary is the applied power. Higher applied power will increase the net signal intensity but not necessarily improve the detection limit. The following information may prove useful:
IMPORTANT: Sample Ar Gas Flow cannot be separated from Applied Power and Sampling Depth for ICP-MS. The sample Ar gas flow for ICP-MS systems is a parameter that is more complex than with ICP-OES instrumentation. Assuming the goal is to obtain the maximum signal intensity, the Ar gas flow is closely related to the applied power and sampling depth. There is not a single set of optimum power, sampling depth, and sample Ar flow settings. For example, a higher applied power will increase the signal intensity but change the optimum sampling depth and sample Ar flow. However, the higher sample Ar flow rates required at high power bring about some degradation in other performance characteristics. If the applied power is constant for every method, then the optimum sampling depth will change as the sample Ar is changed. The consideration of MO (metal oxide) formation and different sensitivities at different mass ranges must also be made with increased sample Ar flow. Here are some final observations that may prove useful:
The above observations may seem confusing, but in reality they give the operator a degree of flexibility that the ICP-OES operator does not have in that you can optimize the instrument for selected mass ranges. For example, we know that a higher applied power will increase the signal intensity. We also know that there is an optimum gas flow for each nebulizer. Therefore, if an applied power of 1.35 kW is selected and we know that our nebulizer performs best at an Ar gas flow of 1.0 L/min then the nest step is to adjust the sampling depth to give the optimum signal while aspirating a solution containing a combination of light, mid-range, and heavy elements such as Mg, Rh, Ce, and U. If the double ion or MO signals are higher than desirable, a reduction in the peristaltic pump tubing diameter or pumping speed should lower these signals. These initial adjustments will take a lot of time and patience but they are well worth the effort. As the operator makes adjustments in these key parameters, a pattern will begin to unfold allowing the operator to optimize the instrument for selected mass ranges. It is suggested that new ICP-MS operators take the time to determine the trends when changes in applied power, sample Ar flow, sampling depth and peristaltic pump speed are made. Further Reading |
|
Contact Inorganic Ventures for all your inorganic standard needs:
1.800.669.6799 ♦ 1.540.585.3030 info@inorganicventures.com |
|
| INORGANIC VENTURES and WE FLEX TO YOUR SPECS are registered trademarks of Inorganic Ventures, Inc. ©1985-2010 Inorganic Ventures, Inc. All rights reserved. Privacy | Terms |
|