Eliminating High and Erratic Blanks

Guillaume contacted Inorganic Ventures to find out whether we knew of a published paper with the values of Environmental Contamination: Table 8.1, Effect of Laboratory Atmosphere on Pb Blank Levels. He noted that his lab was having some problems with the lead blanks values (ICP-MS), which appeared random to him.

Dear Guillaume,

We used to use leaded gasoline in the USA, making ppb level Pb determinations nearly impossible. The basic rule back then was to work in the hood only if you must as it was the greatest area for contamination because of increased air exposure, but always keep the sample covered. Most contamination today is from apparatus and reagents.

I would like to suggest that you perform blanks in and away from the hood and covered and uncovered, with varying amounts of reagents etc. in order to identify the source of contamination (air, apparatus, operator, or reagents). All four areas come into play with trace Pb determinations and each laboratory must perform their personal studies to identify, and hopefully eliminate, high and erratic blanks.

Thank you for contacting Inorganic Ventures and for your question.

Serving you in chemistry,
Paul R. Gaines Paul R. Gaines, Ph.D.
CEO of Inorganic Ventures & Fellow Chemist

DISCLAIMER: Advice offered by the chemists at Inorganic Ventures is intended for the individual posing the question. Feel free to contact us to verify whether these suggestions apply to your unique circumstances.

Further Reading

Have a question?

Share your analytical chemistry problem with our experienced team of scientists, including well-known spectroscopist Dr. Paul Gaines. Note that while we make an effort to respond to every question, customers of Inorganic Ventures are always given priority.

Follow us on TwitterBecome a fan on FacebookSubscribe to our feed

Fully ISO Accredited

Get the IV Line eNewsletter!

Custom Standards by Inorganic Ventures

Request an Inorganic Ventures catalog
Inorganic Ventures spectrum